技術參數
Dimensions | 1200 | 1500 | 1800 |
---|---|---|---|
Width [mm] | 1200 | 1500 | 1800 |
Depth [mm] | 900 | ||
Height [mm] | 2700 | ||
Clear width, intern al workspace [mm] | 1150 | 1450 | 1750 |
Clear heigl1t, intern al workspace [mm] | 1060 | ||
Working height [mm ] | 900 | ||
Filter housing, width x depth x height [mm] | 820 X 775 X 674 |
Weight | 1200 | 1500 | 1800 |
---|---|---|---|
Without installations and fume-scrubber [kg] |
Approx. 250
Approx. 300
|
||
Filter housing [kg] | 90 |
Material/surface | |
---|---|
Worklop |
Stoneware Polypropylene Epoxy Stainless steel |
Internal lining |
Melamine resin facing Solid grade laminate |
Ventilation technology | 1200 | 1500 | 1500 |
---|---|---|---|
Minimum air exchange rate [m3/h](1) | 480 | 600 | 720 |
Pressure loss, filter [Pa](2) | 35/200 | 45/235 | 65/235 |
Pressure loss, particle filter [Pa](2) | 70/300 | 45/365 | 130/430 |
Function display | FAZ | ||
Airflow damper, constant | Airflow-Controller AC | ||
Airflow damper, variable | Airflow-Controller AC | ||
Detector of sash posit ion | Only variable with Airflow -Controller AC |
註解說明:
- (1) 所有排風量指標均為面窗開啟500mm高度時(根據EN14175)測試的資料,跟蹤氣體最大值為德國標準(BG RCI )推薦的數值
- (2) 壓力損失值是指清潔/污染的狀態。
抽風櫃氣流閥的最大進氣壓力值不應超過600 Pa。
標示的最低換氣率根據EN14175-3規定的測試條件而定。設計通風系統大小時,必須滿足此最低換氣率。
若現場使用了排風監測系統或者氣流閥,排風量要求會有所不同。運行條件必須徵得Waldner同意。
對於帶有過濾器的通風櫃,必須將集成過濾器級的壓力損失加到通風櫃的壓力損失上。